Tying a loop-to-strand suture: is it safe?

Am J Obstet Gynecol. 2005 Apr;192(4):1094-7. doi: 10.1016/j.ajog.2004.09.032.

Abstract

Objective: This study was undertaken to evaluate the integrity of the loop-to-strand knot when tied with square and nonidentical sliding knots.

Study design: The synthetic absorbable monofilament suture poliglecaprone 25 in 0 and 2-0 suture gauges was used in this experiment. For each suture gauge, 3 groups of knots were tested: (1) single strand-to-single strand, flat square knot, (2) loop-to-single strand, flat square knot, and (3) loop-to-single strand, nonidentical sliding knot. All knots were tied with 6 throws. The proportion of knots becoming untied was compared among the 3 groups for each suture gauge. Ultimate load required to untie or break knots within each group was also evaluated.

Results: The loop-to-strand knot performed well in both suture gauges tested as long as it was tied with a flat square knot. The loop-to-strand knot tied with a nonidentical sliding knot had an unacceptably high failure rate.

Conclusion: The loop-to-strand termination of a continuous suture may be acceptable when tied with a 6-throw flat square knot but not acceptable if tied with sliding knots.

Publication types

  • Comparative Study

MeSH terms

  • Catgut
  • Dioxanes / chemistry*
  • Female
  • Gynecologic Surgical Procedures / instrumentation
  • Gynecologic Surgical Procedures / methods
  • Humans
  • Materials Testing
  • Polyesters / chemistry*
  • Safety
  • Sensitivity and Specificity
  • Suture Techniques*
  • Sutures
  • Tensile Strength

Substances

  • Dioxanes
  • Polyesters
  • glycolide E-caprolactone copolymer